File name: Pandora83s1_BondvilleIL_20241105_L2Fit_ffus5c2d20230316p1-8.txt
File generation date: 20241106T083020.4Z
Data description: Level 2 spectral fitting results file
Data file version: ffus5c2p1-8
Local principal investigator: Alexander Cede
Network principal investigator: Alexander Cede
Instrument type: Pandora
Instrument number: 83
Spectrometer number: 1
Processing software version used: BlickP v1.8.60
Instrument operation file used: Pandora83_OF_v3d20231115.txt
Instrument calibration file used: Pandora83s1_CF_v2d20240316.txt
Level 1 file used: Pandora83s1_BondvilleIL_20241105_L1_smca1c2d20230316p1-8.txt
Full location name: Bondville, Illinois EPA Castnet Site
Short location name: BondvilleIL
Country of location: USA
Location latitude [deg]: 40.0520
Location longitude [deg]: -88.3725
Location altitude [m]: 213
Local noon date: 20241105
Notes on s-number (L1 configuration): Corrections NOT applied although requested by the s-number are latency correction, matrix method stray light correction (replaced by simple method)
Data caveats: OOB session based on clearance session 1
First and last pixel inside fitting window: 324 590
Nominal wavelengths inside fitting window [nm]: 322.5539 322.6932 322.8323 322.9715 323.1107 323.2499 323.3890 323.5281 323.6673 323.8064 323.9455 324.0846 324.2236 324.3627 324.5018 324.6408 324.7798 324.9188 325.0578 325.1968 325.3358 325.4748 325.6138 325.7527 325.8916 326.0306 326.1695 326.3084 326.4473 326.5861 326.7250 326.8639 327.0027 327.1415 327.2803 327.4192 327.5580 327.6967 327.8355 327.9743 328.1130 328.2517 328.3905 328.5292 328.6679 328.8066 328.9453 329.0839 329.2226 329.3612 329.4998 329.6385 329.7771 329.9157 330.0542 330.1928 330.3314 330.4699 330.6085 330.7470 330.8855 331.0240 331.1625 331.3010 331.4394 331.5779 331.7163 331.8547 331.9932 332.1316 332.2700 332.4083 332.5467 332.6851 332.8234 332.9618 333.1001 333.2384 333.3767 333.5150 333.6532 333.7915 333.9298 334.0680 334.2062 334.3444 334.4826 334.6208 334.7590 334.8972 335.0353 335.1735 335.3116 335.4497 335.5878 335.7259 335.8640 336.0021 336.1401 336.2782 336.4162 336.5543 336.6923 336.8303 336.9683 337.1062 337.2442 337.3821 337.5201 337.6580 337.7959 337.9338 338.0717 338.2096 338.3475 338.4853 338.6232 338.7610 338.8988 339.0366 339.1744 339.3122 339.4500 339.5877 339.7255 339.8632 340.0010 340.1387 340.2764 340.4140 340.5517 340.6894 340.8270 340.9647 341.1023 341.2399 341.3775 341.5151 341.6527 341.7903 341.9278 342.0653 342.2029 342.3404 342.4779 342.6154 342.7529 342.8903 343.0278 343.1652 343.3027 343.4401 343.5775 343.7149 343.8523 343.9897 344.1270 344.2644 344.4017 344.5390 344.6763 344.8136 344.9509 345.0882 345.2254 345.3627 345.4999 345.6372 345.7744 345.9116 346.0488 346.1859 346.3231 346.4603 346.5974 346.7345 346.8716 347.0087 347.1458 347.2829 347.4200 347.5570 347.6941 347.8311 347.9681 348.1051 348.2421 348.3791 348.5160 348.6530 348.7899 348.9269 349.0638 349.2007 349.3376 349.4745 349.6113 349.7482 349.8850 350.0219 350.1587 350.2955 350.4323 350.5691 350.7058 350.8426 350.9793 351.1160 351.2528 351.3895 351.5262 351.6628 351.7995 351.9362 352.0728 352.2094 352.3461 352.4827 352.6193 352.7558 352.8924 353.0290 353.1655 353.3020 353.4386 353.5751 353.7116 353.8480 353.9845 354.1210 354.2574 354.3938 354.5303 354.6667 354.8031 354.9394 355.0758 355.2122 355.3485 355.4848 355.6212 355.7575 355.8938 356.0300 356.1663 356.3026 356.4388 356.5750 356.7112 356.8474 356.9836 357.1198 357.2560 357.3921 357.5283 357.6644 357.8005 357.9366 358.0727 358.2088 358.3448 358.4809 358.6169 358.7530 358.8890 359.0250 359.1610
---------------------------------------------------------------------------------------
Column 1: Two letter code of measurement routine
Column 2: UT date and time for center-time of measurement, yyyymmddThhmmssZ (ISO 8601)
Column 3: Fractional days since 1-Jan-2000 midnight for center-time of measurement
Column 4: Routine count (1 for the first routine of the day, 2 for the second, etc.)
Column 5: Repetition count (1 for the first set in the routine, 2 for the second, etc.)
Column 6: Total duration of measurement set in seconds
Column 7: Data processing type index
Column 8: Solar zenith angle for center-time of measurement in degree
Column 9: Solar azimuth for center-time of measurement in degree, 0=north, increases clockwise
Column 10: Lunar zenith angle for center-time of measurement in degree
Column 11: Lunar azimuth for center-time of measurement in degree, 0=north, increases clockwise
Column 12: Pointing zenith angle in degree, absolute or relative (see next column), 999=tracker not used
Column 13: Zenith pointing mode: zenith angle is... 0=absolute, 1=relative to sun, 2=relative to moon
Column 14: Pointing azimuth in degree, increases clockwise, absolute (0=north) or relative (see next column), 999=tracker not used
Column 15: Azimuth pointing mode: like zenith angle mode but also fixed scattering angles relative to sun (3) or moon (4)
Column 16: Fitting result index: 0=no error or warning, 1, 2=warning, >2=error
Column 17: Number of function evaluations used, 0=linear fitting or fitting not successful or no fitting done
Column 18: rms of unweighted fitting residuals, -9=fitting not successful
Column 19: Normalized rms of fitting residuals weighted with independent uncertainty, -9=fitting not successful or no uncertainty used
Column 20: Expected rms based on independent uncertainty, -9=fitting not successful or no uncertainty given
Column 21: Expected normalized weighted rms based on independent uncertainty, -9=fitting not successful or no uncertainty given
Column 22: Bromine oxide slant column amount [moles per square meter], -9e99=fitting not successful
Column 23: Independent uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 24: Structured uncertainty of bromine oxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 25: Common uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 26: rms-based uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 27: Bromine oxide effective temperature [K]
Column 28: Independent uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 29: Structured uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 30: Common uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 31: Effective bromine oxide fitting wavelength [nm], -9=fitting not successful
Column 32: Diffuse correction applied before fitting at effective fitting wavelength for bromine oxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 33: Formaldehyde slant column amount [moles per square meter], -9e99=fitting not successful
Column 34: Independent uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 35: Structured uncertainty of formaldehyde slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 36: Common uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 37: rms-based uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 38: Formaldehyde effective temperature [K]
Column 39: Independent uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 40: Structured uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 41: Common uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 42: Effective formaldehyde fitting wavelength [nm], -9=fitting not successful
Column 43: Diffuse correction applied before fitting at effective fitting wavelength for formaldehyde [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 44: Nitrous acid slant column amount [moles per square meter], -9e99=fitting not successful
Column 45: Independent uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 46: Structured uncertainty of nitrous acid slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 47: Common uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 48: rms-based uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 49: Nitrous acid effective temperature [K]
Column 50: Independent uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 51: Structured uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 52: Common uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 53: Effective nitrous acid fitting wavelength [nm], -9=fitting not successful
Column 54: Diffuse correction applied before fitting at effective fitting wavelength for nitrous acid [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 55: Nitrogen dioxide slant column amount [moles per square meter], -9e99=fitting not successful
Column 56: Independent uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 57: Structured uncertainty of nitrogen dioxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 58: Common uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 59: rms-based uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 60: Nitrogen dioxide effective temperature [K]
Column 61: Independent uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 62: Structured uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 63: Common uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 64: Effective nitrogen dioxide fitting wavelength [nm], -9=fitting not successful
Column 65: Diffuse correction applied before fitting at effective fitting wavelength for nitrogen dioxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 66: Oxygen dimer slant column amount [moles squared per meter to the 5th], -9e99=fitting not successful
Column 67: Independent uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 68: Structured uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 69: Common uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 70: rms-based uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 71: Oxygen dimer effective temperature [K]
Column 72: Independent uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 73: Structured uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 74: Common uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 75: Effective oxygen dimer fitting wavelength [nm], -9=fitting not successful
Column 76: Diffuse correction applied before fitting at effective fitting wavelength for oxygen dimer [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 77: Ozone slant column amount [moles per square meter], -9e99=fitting not successful
Column 78: Independent uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 79: Structured uncertainty of ozone slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 80: Common uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 81: rms-based uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 82: Ozone effective temperature [K]
Column 83: Independent uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 84: Structured uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 85: Common uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 86: Effective ozone fitting wavelength [nm], -9=fitting not successful
Column 87: Diffuse correction applied before fitting at effective fitting wavelength for ozone [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 88: Sulfur dioxide slant column amount [moles per square meter], -9e99=fitting not successful
Column 89: Independent uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 90: Structured uncertainty of sulfur dioxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 91: Common uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 92: rms-based uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 93: Sulfur dioxide effective temperature [K]
Column 94: Independent uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 95: Structured uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 96: Common uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 97: Effective sulfur dioxide fitting wavelength [nm], -9=fitting not successful
Column 98: Diffuse correction applied before fitting at effective fitting wavelength for sulfur dioxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 99: Molecular scattering air mass factor used for molecular scattering subtraction before the fitting
Column 100: Estimated uncertainty of molecular scattering air mass factor, -9=molecular scattering was not subtracted before the fitting
Column 101: Lower limit used for wavelength scaling [nm]
Column 102: Upper limit used for wavelength scaling [nm]
Column 103: Order of smoothing polynomial
Column 104: Smoothing polynomial coefficient, order 0
Column 105: Independent uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 106: Structured uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 107: Common uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 108: rms-based uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 109: Smoothing polynomial coefficient, order 1
Column 110: Independent uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 111: Structured uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 112: Common uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 113: rms-based uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 114: Smoothing polynomial coefficient, order 2
Column 115: Independent uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 116: Structured uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 117: Common uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 118: rms-based uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 119: Smoothing polynomial coefficient, order 3
Column 120: Independent uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 121: Structured uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 122: Common uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 123: rms-based uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 124: Smoothing polynomial coefficient, order 4
Column 125: Independent uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 126: Structured uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 127: Common uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 128: rms-based uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 129: Mean value of measured data inside fitting window [same units as measurements]
Column 130: Offset polynomial coefficient, order 0
Column 131: Independent uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 132: Structured uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 133: Common uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 134: rms-based uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 135: Wavelength change polynomial coefficient, order 0
Column 136: Independent uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 137: Structured uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 138: Common uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 139: rms-based uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 140: L1 based wavelength change coefficient, order 0, -9=no L1 wavelength change determination
Column 141: L1 based wavelength change coefficient, order 1, -9=no L1 wavelength change determination
Column 142: Resolution change polynomial coefficient, order 0
Column 143: Independent uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 144: Structured uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 145: Common uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 146: rms-based uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 147: L2Fit data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality
Column 148: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ1 limit, 0=L1 data quality above 0, 1=Spectral fitting was not successful, 2=Wavelength shift too large, 3=Normalized rms of fitting residuals weighted with independent uncertainty too large
Column 149: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ2 limit (same parameters as for DQ1)
Column 150: L1 data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality
Column 151: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ1 limit, 0=Saturated data, 1=Too few dark counts measurements, 2=No temperature given or effective temperature too different from the reference temperature, 3=Dark count too high, 4=Unsuccessful dark background fitting, 5=The dark count differs significantly from the dark map for too many pixels, 6=Absolute value of estimated average residual stray light level too high, 7=Although attempted, no wavelength change could be retrieved, 8=Absolute value of retrieved wavelength shift too large, 9=Retrieved wavelength shift differs too much from the shift predicted by the effective temperature
Column 152: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ2 limit (same parameters as for DQ1)
Column 153: Atmospheric variability [%], 999=no atmospheric variability was determined
Column 154: Wavelength effective temperature [�C], 999=no effective temperature given
Column 155: Estimated average residual stray light level [%] (only valid for stray light correction methods 2 and higher)
Column 156: Retrieved wavelength shift from L1 data [nm], -9=no wavelength change determination
Column 157: Retrieved total wavelength shift [nm], -9=no wavelength change fitting
Column 158: Number of bright count cycles
Column 159: Number of dark count cycles
Column 160: Effective position of filterwheel #1, 0=filterwheel not used, 1-9 are valid positions
Column 161: Effective position of filterwheel #2, 0=filterwheel not used, 1-9 are valid positions
Column 162: Sum over 2^i, 0=spectra were interpolated in time, 1=spectra are corrected for off-target signal
Column 163: Integration time [ms]
Column 164: Temperature at electronics board [�C], 999=no temperature signal
Column 165: Spectrometer control temperature [�C], 999=no temperature signal
Column 166: Auxiliary spectrometer temperature [�C], 999=no temperature signal
Column 167: Temperature in head sensor [�C], 999=no temperature signal
Column 168: Humidity in head sensor [%], -9=no humidity signal
Column 169: Pressure in head sensor [hPa], -9=no pressure signal
Columns 170-436: Unweighted slant column residuals for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting
Columns 437-703: Normalized slant column residuals weighted with independent instrumental uncertainty for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting, 0=no independent instrumental uncertainty was given
---------------------------------------------------------------------------------------