File name: Pandora229s1_Daegu_20250120_L2Fit_ffus5c4d20221031p1-8.txt
File generation date: 20250221T212608.7Z
Data description: Level 2 spectral fitting results file
Data file version: ffus5c4p1-8
Local principal investigator: Sunyoung Park
Network principal investigator: Alexander Cede
Instrument type: Pandora
Instrument number: 229
Spectrometer number: 1
Processing software version used: BlickP v1.8.74
Instrument operation file used: Pandora229_OF_v3d20220728.txt
Instrument calibration file used: Pandora229s1_CF_v4d20221031.txt
Level 1 file used: Pandora229s1_Daegu_20250120_L1_smca1c4d20221031p1-8.txt
Full location name: Kyungpook National University
Short location name: Daegu
Country of location: Korea, Republic of
Location latitude [deg]: 35.8869
Location longitude [deg]: 128.6061
Location altitude [m]: 44
Local noon date: 20250120
Notes on s-number (L1 configuration): Corrections NOT applied although requested by the s-number are latency correction, matrix method stray light correction (replaced by simple method)
Data caveats: Based on session 3 adding wavelength change parameters to OoB
First and last pixel inside fitting window: 322 586
Nominal wavelengths inside fitting window [nm]: 322.5944 322.7342 322.8740 323.0138 323.1535 323.2933 323.4330 323.5727 323.7125 323.8522 323.9919 324.1315 324.2712 324.4109 324.5505 324.6901 324.8298 324.9694 325.1090 325.2486 325.3881 325.5277 325.6673 325.8068 325.9463 326.0858 326.2253 326.3648 326.5043 326.6438 326.7832 326.9227 327.0621 327.2015 327.3409 327.4803 327.6197 327.7591 327.8985 328.0378 328.1771 328.3165 328.4558 328.5951 328.7344 328.8736 329.0129 329.1522 329.2914 329.4306 329.5698 329.7090 329.8482 329.9874 330.1266 330.2657 330.4049 330.5440 330.6831 330.8222 330.9613 331.1004 331.2395 331.3786 331.5176 331.6566 331.7957 331.9347 332.0737 332.2127 332.3516 332.4906 332.6296 332.7685 332.9074 333.0463 333.1852 333.3241 333.4630 333.6019 333.7407 333.8796 334.0184 334.1572 334.2960 334.4348 334.5736 334.7124 334.8511 334.9899 335.1286 335.2673 335.4060 335.5447 335.6834 335.8221 335.9607 336.0994 336.2380 336.3766 336.5152 336.6538 336.7924 336.9310 337.0695 337.2081 337.3466 337.4851 337.6237 337.7622 337.9006 338.0391 338.1776 338.3160 338.4545 338.5929 338.7313 338.8697 339.0081 339.1465 339.2848 339.4232 339.5615 339.6998 339.8381 339.9764 340.1147 340.2530 340.3913 340.5295 340.6678 340.8060 340.9442 341.0824 341.2206 341.3588 341.4969 341.6351 341.7732 341.9114 342.0495 342.1876 342.3257 342.4637 342.6018 342.7399 342.8779 343.0159 343.1539 343.2919 343.4299 343.5679 343.7059 343.8438 343.9818 344.1197 344.2576 344.3955 344.5334 344.6713 344.8091 344.9470 345.0848 345.2227 345.3605 345.4983 345.6361 345.7738 345.9116 346.0493 346.1871 346.3248 346.4625 346.6002 346.7379 346.8756 347.0133 347.1509 347.2885 347.4262 347.5638 347.7014 347.8390 347.9765 348.1141 348.2517 348.3892 348.5267 348.6642 348.8017 348.9392 349.0767 349.2142 349.3516 349.4890 349.6265 349.7639 349.9013 350.0387 350.1760 350.3134 350.4507 350.5881 350.7254 350.8627 351.0000 351.1373 351.2745 351.4118 351.5491 351.6863 351.8235 351.9607 352.0979 352.2351 352.3723 352.5094 352.6466 352.7837 352.9208 353.0579 353.1950 353.3321 353.4691 353.6062 353.7432 353.8803 354.0173 354.1543 354.2913 354.4283 354.5652 354.7022 354.8391 354.9760 355.1130 355.2499 355.3867 355.5236 355.6605 355.7973 355.9342 356.0710 356.2078 356.3446 356.4814 356.6182 356.7549 356.8917 357.0284 357.1651 357.3018 357.4385 357.5752 357.7119 357.8485 357.9852 358.1218 358.2584 358.3950 358.5316 358.6682 358.8048 358.9413 359.0779
---------------------------------------------------------------------------------------
Column 1: Two letter code of measurement routine
Column 2: UT date and time for center-time of measurement, yyyymmddThhmmssZ (ISO 8601)
Column 3: Fractional days since 1-Jan-2000 midnight for center-time of measurement
Column 4: Routine count (1 for the first routine of the day, 2 for the second, etc.)
Column 5: Repetition count (1 for the first set in the routine, 2 for the second, etc.)
Column 6: Total duration of measurement set in seconds
Column 7: Data processing type index
Column 8: Solar zenith angle for center-time of measurement in degree
Column 9: Solar azimuth for center-time of measurement in degree, 0=north, increases clockwise
Column 10: Lunar zenith angle for center-time of measurement in degree
Column 11: Lunar azimuth for center-time of measurement in degree, 0=north, increases clockwise
Column 12: Pointing zenith angle in degree, absolute or relative (see next column), 999=tracker not used
Column 13: Zenith pointing mode: zenith angle is... 0=absolute, 1=relative to sun, 2=relative to moon
Column 14: Pointing azimuth in degree, increases clockwise, absolute (0=north) or relative (see next column), 999=tracker not used
Column 15: Azimuth pointing mode: like zenith angle mode but also fixed scattering angles relative to sun (3) or moon (4)
Column 16: Fitting result index: 0=no error or warning, 1, 2=warning, >2=error
Column 17: Number of function evaluations used, 0=linear fitting or fitting not successful or no fitting done
Column 18: rms of unweighted fitting residuals, -9=fitting not successful
Column 19: Normalized rms of fitting residuals weighted with independent uncertainty, -9=fitting not successful or no uncertainty used
Column 20: Expected rms based on independent uncertainty, -9=fitting not successful or no uncertainty given
Column 21: Expected normalized weighted rms based on independent uncertainty, -9=fitting not successful or no uncertainty given
Column 22: Bromine oxide slant column amount [moles per square meter], -9e99=fitting not successful
Column 23: Independent uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 24: Structured uncertainty of bromine oxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 25: Common uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 26: rms-based uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 27: Bromine oxide effective temperature [K]
Column 28: Independent uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 29: Structured uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 30: Common uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 31: Effective bromine oxide fitting wavelength [nm], -9=fitting not successful
Column 32: Diffuse correction applied before fitting at effective fitting wavelength for bromine oxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 33: Formaldehyde slant column amount [moles per square meter], -9e99=fitting not successful
Column 34: Independent uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 35: Structured uncertainty of formaldehyde slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 36: Common uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 37: rms-based uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 38: Formaldehyde effective temperature [K]
Column 39: Independent uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 40: Structured uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 41: Common uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 42: Effective formaldehyde fitting wavelength [nm], -9=fitting not successful
Column 43: Diffuse correction applied before fitting at effective fitting wavelength for formaldehyde [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 44: Nitrous acid slant column amount [moles per square meter], -9e99=fitting not successful
Column 45: Independent uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 46: Structured uncertainty of nitrous acid slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 47: Common uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 48: rms-based uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 49: Nitrous acid effective temperature [K]
Column 50: Independent uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 51: Structured uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 52: Common uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 53: Effective nitrous acid fitting wavelength [nm], -9=fitting not successful
Column 54: Diffuse correction applied before fitting at effective fitting wavelength for nitrous acid [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 55: Nitrogen dioxide slant column amount [moles per square meter], -9e99=fitting not successful
Column 56: Independent uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 57: Structured uncertainty of nitrogen dioxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 58: Common uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 59: rms-based uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 60: Nitrogen dioxide effective temperature [K]
Column 61: Independent uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 62: Structured uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 63: Common uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 64: Effective nitrogen dioxide fitting wavelength [nm], -9=fitting not successful
Column 65: Diffuse correction applied before fitting at effective fitting wavelength for nitrogen dioxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 66: Oxygen dimer slant column amount [moles squared per meter to the 5th], -9e99=fitting not successful
Column 67: Independent uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 68: Structured uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 69: Common uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 70: rms-based uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 71: Oxygen dimer effective temperature [K]
Column 72: Independent uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 73: Structured uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 74: Common uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 75: Effective oxygen dimer fitting wavelength [nm], -9=fitting not successful
Column 76: Diffuse correction applied before fitting at effective fitting wavelength for oxygen dimer [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 77: Ozone slant column amount [moles per square meter], -9e99=fitting not successful
Column 78: Independent uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 79: Structured uncertainty of ozone slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 80: Common uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 81: rms-based uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 82: Ozone effective temperature [K]
Column 83: Independent uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 84: Structured uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 85: Common uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 86: Effective ozone fitting wavelength [nm], -9=fitting not successful
Column 87: Diffuse correction applied before fitting at effective fitting wavelength for ozone [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 88: Sulfur dioxide slant column amount [moles per square meter], -9e99=fitting not successful
Column 89: Independent uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 90: Structured uncertainty of sulfur dioxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 91: Common uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful
Column 92: rms-based uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful
Column 93: Sulfur dioxide effective temperature [K]
Column 94: Independent uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful
Column 95: Structured uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful
Column 96: Common uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful
Column 97: Effective sulfur dioxide fitting wavelength [nm], -9=fitting not successful
Column 98: Diffuse correction applied before fitting at effective fitting wavelength for sulfur dioxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction
Column 99: Molecular scattering air mass factor used for molecular scattering subtraction before the fitting
Column 100: Estimated uncertainty of molecular scattering air mass factor, -9=molecular scattering was not subtracted before the fitting
Column 101: Lower limit used for wavelength scaling [nm]
Column 102: Upper limit used for wavelength scaling [nm]
Column 103: Order of smoothing polynomial
Column 104: Smoothing polynomial coefficient, order 0
Column 105: Independent uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 106: Structured uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 107: Common uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 108: rms-based uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful
Column 109: Smoothing polynomial coefficient, order 1
Column 110: Independent uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 111: Structured uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 112: Common uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 113: rms-based uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful
Column 114: Smoothing polynomial coefficient, order 2
Column 115: Independent uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 116: Structured uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 117: Common uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 118: rms-based uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful
Column 119: Smoothing polynomial coefficient, order 3
Column 120: Independent uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 121: Structured uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 122: Common uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 123: rms-based uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful
Column 124: Smoothing polynomial coefficient, order 4
Column 125: Independent uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 126: Structured uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 127: Common uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 128: rms-based uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful
Column 129: Mean value of measured data inside fitting window [same units as measurements]
Column 130: Offset polynomial coefficient, order 0
Column 131: Independent uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 132: Structured uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 133: Common uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 134: rms-based uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful
Column 135: Wavelength change polynomial coefficient, order 0
Column 136: Independent uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 137: Structured uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 138: Common uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 139: rms-based uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful
Column 140: L1 based wavelength change coefficient, order 0, -9=no L1 wavelength change determination
Column 141: L1 based wavelength change coefficient, order 1, -9=no L1 wavelength change determination
Column 142: Resolution change polynomial coefficient, order 0
Column 143: Independent uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 144: Structured uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 145: Common uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 146: rms-based uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful
Column 147: L2Fit data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality
Column 148: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ1 limit, 0=L1 data quality above 0, 1=Spectral fitting was not successful, 2=Wavelength shift too large, 3=Normalized rms of fitting residuals weighted with independent uncertainty too large
Column 149: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ2 limit (same parameters as for DQ1)
Column 150: L1 data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality
Column 151: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ1 limit, 0=Saturated data, 1=Too few dark counts measurements, 2=No temperature given or effective temperature too different from the reference temperature, 3=Dark count too high, 4=Unsuccessful dark background fitting, 5=The dark count differs significantly from the dark map for too many pixels, 6=Absolute value of estimated average residual stray light level too high, 7=Although attempted, no wavelength change could be retrieved, 8=Absolute value of retrieved wavelength shift too large, 9=Retrieved wavelength shift differs too much from the shift predicted by the effective temperature
Column 152: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ2 limit (same parameters as for DQ1)
Column 153: Atmospheric variability [%], 999=no atmospheric variability was determined
Column 154: Wavelength effective temperature [�C], 999=no effective temperature given
Column 155: Estimated average residual stray light level [%] (only valid for stray light correction methods 2 and higher)
Column 156: Retrieved wavelength shift from L1 data [nm], -9=no wavelength change determination
Column 157: Retrieved total wavelength shift [nm], -9=no wavelength change fitting
Column 158: Number of bright count cycles
Column 159: Number of dark count cycles
Column 160: Effective position of filterwheel #1, 0=filterwheel not used, 1-9 are valid positions
Column 161: Effective position of filterwheel #2, 0=filterwheel not used, 1-9 are valid positions
Column 162: Sum over 2^i, 0=spectra were interpolated in time, 1=spectra are corrected for off-target signal
Column 163: Integration time [ms]
Column 164: Temperature at electronics board [�C], 999=no temperature signal
Column 165: Spectrometer control temperature [�C], 999=no temperature signal
Column 166: Auxiliary spectrometer temperature [�C], 999=no temperature signal
Column 167: Temperature in head sensor [�C], 999=no temperature signal
Column 168: Humidity in head sensor [%], -9=no humidity signal
Column 169: Pressure in head sensor [hPa], -9=no pressure signal
Columns 170-434: Unweighted slant column residuals for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting
Columns 435-699: Normalized slant column residuals weighted with independent instrumental uncertainty for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting, 0=no independent instrumental uncertainty was given
---------------------------------------------------------------------------------------