File name: Pandora145s1_Toronto-Scarborough_20231207_L2Fit_ffus5c9d20230821p1-8.txt File generation date: 20241123T061220.7Z Data description: Level 2 spectral fitting results file Data file version: ffus5c9p1-8 Local principal investigator: Vitali Fioletov Network principal investigator: Alexander Cede Instrument type: Pandora Instrument number: 145 Spectrometer number: 1 Processing software version used: BlickP v1.8.60 Instrument operation file used: Pandora145_OF_v5d20190718.txt Instrument calibration file used: Pandora145s1_CF_v9d20230821.txt Level 1 file used: Pandora145s1_Toronto-Scarborough_20231207_L1_smca1c9d20230821p1-8.txt Full location name: University of Toronto - Scarborough Short location name: Toronto-Scarborough Country of location: Canada Location latitude [deg]: 43.7843 Location longitude [deg]: -79.1874 Location altitude [m]: 137 Local noon date: 20231207 Notes on s-number (L1 configuration): Corrections NOT applied although requested by the s-number are dark map for dark correction (replaced by immediate dark method), latency correction, matrix method stray light correction (replaced by simple method) Data caveats: field cal for autumn2023 ff, based on Session 8(7) First and last pixel inside fitting window: 351 616 Nominal wavelengths inside fitting window [nm]: 322.5625 322.7019 322.8413 322.9807 323.1200 323.2594 323.3988 323.5381 323.6774 323.8167 323.9561 324.0954 324.2346 324.3739 324.5132 324.6524 324.7917 324.9309 325.0701 325.2093 325.3485 325.4877 325.6268 325.7660 325.9051 326.0443 326.1834 326.3225 326.4616 326.6007 326.7398 326.8788 327.0179 327.1569 327.2960 327.4350 327.5740 327.7130 327.8520 327.9909 328.1299 328.2688 328.4078 328.5467 328.6856 328.8245 328.9634 329.1023 329.2411 329.3800 329.5188 329.6577 329.7965 329.9353 330.0741 330.2129 330.3516 330.4904 330.6291 330.7679 330.9066 331.0453 331.1840 331.3227 331.4614 331.6000 331.7387 331.8773 332.0159 332.1546 332.2932 332.4318 332.5703 332.7089 332.8475 332.9860 333.1246 333.2631 333.4016 333.5401 333.6786 333.8170 333.9555 334.0940 334.2324 334.3708 334.5092 334.6476 334.7860 334.9244 335.0628 335.2011 335.3395 335.4778 335.6161 335.7544 335.8927 336.0310 336.1693 336.3075 336.4458 336.5840 336.7223 336.8605 336.9987 337.1369 337.2750 337.4132 337.5514 337.6895 337.8276 337.9657 338.1038 338.2419 338.3800 338.5181 338.6561 338.7942 338.9322 339.0702 339.2082 339.3462 339.4842 339.6222 339.7602 339.8981 340.0360 340.1740 340.3119 340.4498 340.5877 340.7255 340.8634 341.0012 341.1391 341.2769 341.4147 341.5525 341.6903 341.8281 341.9659 342.1036 342.2414 342.3791 342.5168 342.6545 342.7922 342.9299 343.0676 343.2052 343.3429 343.4805 343.6181 343.7557 343.8933 344.0309 344.1685 344.3060 344.4436 344.5811 344.7186 344.8561 344.9936 345.1311 345.2686 345.4061 345.5435 345.6809 345.8184 345.9558 346.0932 346.2306 346.3679 346.5053 346.6427 346.7800 346.9173 347.0546 347.1919 347.3292 347.4665 347.6038 347.7410 347.8783 348.0155 348.1527 348.2899 348.4271 348.5643 348.7014 348.8386 348.9757 349.1129 349.2500 349.3871 349.5242 349.6613 349.7983 349.9354 350.0724 350.2095 350.3465 350.4835 350.6205 350.7574 350.8944 351.0314 351.1683 351.3052 351.4422 351.5791 351.7160 351.8528 351.9897 352.1266 352.2634 352.4002 352.5371 352.6739 352.8107 352.9474 353.0842 353.2210 353.3577 353.4944 353.6312 353.7679 353.9046 354.0412 354.1779 354.3146 354.4512 354.5878 354.7245 354.8611 354.9977 355.1342 355.2708 355.4074 355.5439 355.6804 355.8170 355.9535 356.0900 356.2264 356.3629 356.4994 356.6358 356.7722 356.9087 357.0451 357.1814 357.3178 357.4542 357.5905 357.7269 357.8632 357.9995 358.1358 358.2721 358.4084 358.5447 358.6809 358.8172 358.9534 359.0896 --------------------------------------------------------------------------------------- Column 1: Two letter code of measurement routine Column 2: UT date and time for center-time of measurement, yyyymmddThhmmssZ (ISO 8601) Column 3: Fractional days since 1-Jan-2000 midnight for center-time of measurement Column 4: Routine count (1 for the first routine of the day, 2 for the second, etc.) Column 5: Repetition count (1 for the first set in the routine, 2 for the second, etc.) Column 6: Total duration of measurement set in seconds Column 7: Data processing type index Column 8: Solar zenith angle for center-time of measurement in degree Column 9: Solar azimuth for center-time of measurement in degree, 0=north, increases clockwise Column 10: Lunar zenith angle for center-time of measurement in degree Column 11: Lunar azimuth for center-time of measurement in degree, 0=north, increases clockwise Column 12: Pointing zenith angle in degree, absolute or relative (see next column), 999=tracker not used Column 13: Zenith pointing mode: zenith angle is... 0=absolute, 1=relative to sun, 2=relative to moon Column 14: Pointing azimuth in degree, increases clockwise, absolute (0=north) or relative (see next column), 999=tracker not used Column 15: Azimuth pointing mode: like zenith angle mode but also fixed scattering angles relative to sun (3) or moon (4) Column 16: Fitting result index: 0=no error or warning, 1, 2=warning, >2=error Column 17: Number of function evaluations used, 0=linear fitting or fitting not successful or no fitting done Column 18: rms of unweighted fitting residuals, -9=fitting not successful Column 19: Normalized rms of fitting residuals weighted with independent uncertainty, -9=fitting not successful or no uncertainty used Column 20: Expected rms based on independent uncertainty, -9=fitting not successful or no uncertainty given Column 21: Expected normalized weighted rms based on independent uncertainty, -9=fitting not successful or no uncertainty given Column 22: Bromine oxide slant column amount [moles per square meter], -9e99=fitting not successful Column 23: Independent uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 24: Structured uncertainty of bromine oxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 25: Common uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 26: rms-based uncertainty of bromine oxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 27: Bromine oxide effective temperature [K] Column 28: Independent uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 29: Structured uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 30: Common uncertainty of bromine oxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 31: Effective bromine oxide fitting wavelength [nm], -9=fitting not successful Column 32: Diffuse correction applied before fitting at effective fitting wavelength for bromine oxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 33: Formaldehyde slant column amount [moles per square meter], -9e99=fitting not successful Column 34: Independent uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 35: Structured uncertainty of formaldehyde slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 36: Common uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 37: rms-based uncertainty of formaldehyde slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 38: Formaldehyde effective temperature [K] Column 39: Independent uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 40: Structured uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 41: Common uncertainty of formaldehyde effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 42: Effective formaldehyde fitting wavelength [nm], -9=fitting not successful Column 43: Diffuse correction applied before fitting at effective fitting wavelength for formaldehyde [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 44: Nitrous acid slant column amount [moles per square meter], -9e99=fitting not successful Column 45: Independent uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 46: Structured uncertainty of nitrous acid slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 47: Common uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 48: rms-based uncertainty of nitrous acid slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 49: Nitrous acid effective temperature [K] Column 50: Independent uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 51: Structured uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 52: Common uncertainty of nitrous acid effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 53: Effective nitrous acid fitting wavelength [nm], -9=fitting not successful Column 54: Diffuse correction applied before fitting at effective fitting wavelength for nitrous acid [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 55: Nitrogen dioxide slant column amount [moles per square meter], -9e99=fitting not successful Column 56: Independent uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 57: Structured uncertainty of nitrogen dioxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 58: Common uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 59: rms-based uncertainty of nitrogen dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 60: Nitrogen dioxide effective temperature [K] Column 61: Independent uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 62: Structured uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 63: Common uncertainty of nitrogen dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 64: Effective nitrogen dioxide fitting wavelength [nm], -9=fitting not successful Column 65: Diffuse correction applied before fitting at effective fitting wavelength for nitrogen dioxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 66: Oxygen dimer slant column amount [moles squared per meter to the 5th], -9e99=fitting not successful Column 67: Independent uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 68: Structured uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 69: Common uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 70: rms-based uncertainty of oxygen dimer slant column amount [moles squared per meter to the 5th], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 71: Oxygen dimer effective temperature [K] Column 72: Independent uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 73: Structured uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 74: Common uncertainty of oxygen dimer effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 75: Effective oxygen dimer fitting wavelength [nm], -9=fitting not successful Column 76: Diffuse correction applied before fitting at effective fitting wavelength for oxygen dimer [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 77: Ozone slant column amount [moles per square meter], -9e99=fitting not successful Column 78: Independent uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 79: Structured uncertainty of ozone slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 80: Common uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 81: rms-based uncertainty of ozone slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 82: Ozone effective temperature [K] Column 83: Independent uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 84: Structured uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 85: Common uncertainty of ozone effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 86: Effective ozone fitting wavelength [nm], -9=fitting not successful Column 87: Diffuse correction applied before fitting at effective fitting wavelength for ozone [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 88: Sulfur dioxide slant column amount [moles per square meter], -9e99=fitting not successful Column 89: Independent uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 90: Structured uncertainty of sulfur dioxide slant column amount [moles per square meter], -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 91: Common uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -6=no common uncertainty input was given, -7=not given since method "MEAS" was chosen, -9=spectral fitting not successful Column 92: rms-based uncertainty of sulfur dioxide slant column amount [moles per square meter], -1=cross section is zero in this wavelength range, -3=spectral fitting was done, but no rms-based uncertainty could be retrieved, -9=spectral fitting not successful Column 93: Sulfur dioxide effective temperature [K] Column 94: Independent uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no independent uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -5=no independent uncertainty input was given, -9=spectral fitting not successful Column 95: Structured uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -9=spectral fitting not successful Column 96: Common uncertainty of sulfur dioxide effective temperature [K], -1=temperature fitting was requested, but cross section is zero in this wavelength range, -2=no temperature fitting was requested and output for effective temperature and structured uncertainty of it is based on f-code, -3=spectral fitting was done, but no common uncertainty could be retrieved, -4=temperature fitting was requested, but differential optical depth is too small to retrieve the temperature, -6=no common uncertainty input was given, -9=spectral fitting not successful Column 97: Effective sulfur dioxide fitting wavelength [nm], -9=fitting not successful Column 98: Diffuse correction applied before fitting at effective fitting wavelength for sulfur dioxide [%], 0=no diffuse correction applied or fitting not requested, >0=measured diffuse correction, <0=(negative value of) calculated diffuse correction Column 99: Molecular scattering air mass factor used for molecular scattering subtraction before the fitting Column 100: Estimated uncertainty of molecular scattering air mass factor, -9=molecular scattering was not subtracted before the fitting Column 101: Lower limit used for wavelength scaling [nm] Column 102: Upper limit used for wavelength scaling [nm] Column 103: Order of smoothing polynomial Column 104: Smoothing polynomial coefficient, order 0 Column 105: Independent uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 106: Structured uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 107: Common uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 108: rms-based uncertainty of smoothing polynomial coefficient, order 0, -9=fitting not successful Column 109: Smoothing polynomial coefficient, order 1 Column 110: Independent uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 111: Structured uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 112: Common uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 113: rms-based uncertainty of smoothing polynomial coefficient, order 1, -9=fitting not successful Column 114: Smoothing polynomial coefficient, order 2 Column 115: Independent uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful Column 116: Structured uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful Column 117: Common uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful Column 118: rms-based uncertainty of smoothing polynomial coefficient, order 2, -9=fitting not successful Column 119: Smoothing polynomial coefficient, order 3 Column 120: Independent uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful Column 121: Structured uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful Column 122: Common uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful Column 123: rms-based uncertainty of smoothing polynomial coefficient, order 3, -9=fitting not successful Column 124: Smoothing polynomial coefficient, order 4 Column 125: Independent uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful Column 126: Structured uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful Column 127: Common uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful Column 128: rms-based uncertainty of smoothing polynomial coefficient, order 4, -9=fitting not successful Column 129: Mean value of measured data inside fitting window [same units as measurements] Column 130: Offset polynomial coefficient, order 0 Column 131: Independent uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful Column 132: Structured uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful Column 133: Common uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful Column 134: rms-based uncertainty of offset polynomial coefficient, order 0, -9=fitting not successful Column 135: Wavelength change polynomial coefficient, order 0 Column 136: Independent uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 137: Structured uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 138: Common uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 139: rms-based uncertainty of wavelength change polynomial coefficient, order 0, -9=fitting not successful Column 140: L1 based wavelength change coefficient, order 0, -9=no L1 wavelength change determination Column 141: L1 based wavelength change coefficient, order 1, -9=no L1 wavelength change determination Column 142: Resolution change polynomial coefficient, order 0 Column 143: Independent uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful Column 144: Structured uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful Column 145: Common uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful Column 146: rms-based uncertainty of resolution change polynomial coefficient, order 0, -9=fitting not successful Column 147: L2Fit data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality Column 148: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ1 limit, 0=L1 data quality above 0, 1=Spectral fitting was not successful, 2=Wavelength shift too large, 3=Normalized rms of fitting residuals weighted with independent uncertainty too large Column 149: Sum over 2^i using those i, for which the corresponding L2Fit data quality parameter exceeds the DQ2 limit (same parameters as for DQ1) Column 150: L1 data quality flag: 0=assured high quality, 1=assured medium quality, 2=assured low quality, 10=not-assured high quality, 11=not-assured medium quality, 12=not-assured low quality Column 151: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ1 limit, 0=Saturated data, 1=Too few dark counts measurements, 2=No temperature given or effective temperature too different from the reference temperature, 3=Dark count too high, 4=Unsuccessful dark background fitting, 5=The dark count differs significantly from the dark map for too many pixels, 6=Absolute value of estimated average residual stray light level too high, 7=Although attempted, no wavelength change could be retrieved, 8=Absolute value of retrieved wavelength shift too large, 9=Retrieved wavelength shift differs too much from the shift predicted by the effective temperature Column 152: Sum over 2^i using those i, for which the corresponding L1 data quality parameter exceeds the DQ2 limit (same parameters as for DQ1) Column 153: Atmospheric variability [%], 999=no atmospheric variability was determined Column 154: Wavelength effective temperature [°C], 999=no effective temperature given Column 155: Estimated average residual stray light level [%] (only valid for stray light correction methods 2 and higher) Column 156: Retrieved wavelength shift from L1 data [nm], -9=no wavelength change determination Column 157: Retrieved total wavelength shift [nm], -9=no wavelength change fitting Column 158: Number of bright count cycles Column 159: Number of dark count cycles Column 160: Effective position of filterwheel #1, 0=filterwheel not used, 1-9 are valid positions Column 161: Effective position of filterwheel #2, 0=filterwheel not used, 1-9 are valid positions Column 162: Sum over 2^i, 0=spectra were interpolated in time, 1=spectra are corrected for off-target signal Column 163: Integration time [ms] Column 164: Temperature at electronics board [°C], 999=no temperature signal Column 165: Spectrometer control temperature [°C], 999=no temperature signal Column 166: Auxiliary spectrometer temperature [°C], 999=no temperature signal Column 167: Temperature in head sensor [°C], 999=no temperature signal Column 168: Humidity in head sensor [%], -9=no humidity signal Column 169: Pressure in head sensor [hPa], -9=no pressure signal Columns 170-435: Unweighted slant column residuals for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting Columns 436-701: Normalized slant column residuals weighted with independent instrumental uncertainty for each pixel inside the fitting window multiplied by 1e5, 9e5=pixel was not used for fitting, 0=no independent instrumental uncertainty was given --------------------------------------------------------------------------------------- SS 20231207T164303.6Z 8741.696570 402 1 29.03 2 66.66 173.68 70.09 242.47 0.00 1 0.00 1 0 33 4.066e-03 3.733e-03 3.070e-03 2.874e-03 1.444324e-07 9.154633e-07 1.424043e-07 3.560267e-07 1.270653e-06 222.30 -2.00 3.57 -2.00 339.85 0.00e+00 -3.926026e-04 2.498980e-04 5.377011e-05 3.745389e-05 3.450213e-04 274.93 -2.00 20.00 -2.00 337.37 0.00e+00 -1.097164e-04 3.975595e-05 4.209349e-06 2.382133e-05 6.164419e-05 274.93 -2.00 20.00 -2.00 346.01 0.00e+00 5.459349e-04 1.128967e-04 2.364698e-05 1.534488e-05 1.699910e-04 274.93 -2.00 20.00 -2.00 344.53 0.00e+00 8.330982e+04 2.312324e+05 4.805170e+04 1.852201e+05 3.212560e+05 257.21 -2.00 11.30 -2.00 351.41 0.00e+00 3.169797e-01 3.498179e-03 2.786835e-03 3.086619e-03 4.243680e-03 222.30 -2.00 3.57 -2.00 330.45 0.00e+00 4.970340e-03 4.596634e-03 4.903349e-04 1.012686e-04 4.522976e-03 274.93 -2.00 20.00 -2.00 328.48 0.00e+00 2.505 0.000 319.50 362.20 4 6.4948e+00 3.6575e-03 1.7403e-02 1.3566e-03 5.4520e-03 1.5023e-01 8.9870e-04 2.7252e-03 2.3970e-04 1.2646e-03 -2.9552e-02 1.7784e-03 5.8065e-04 7.2054e-05 2.5011e-03 -9.3149e-04 9.0326e-04 4.7600e-04 2.6141e-05 1.0143e-03 8.3994e-03 1.4073e-03 4.0793e-04 4.9922e-05 1.8013e-03 2.5584e-04 -1.8173e-02 2.1961e-03 2.9140e-04 5.2875e-05 3.1618e-03 2.3767e-05 3.8242e-05 3.2640e-06 4.5364e-07 5.5122e-05 -2.4542e-03 0.0000e+00 7.7752e-01 2.2338e-01 2.0248e-02 2.6645e-03 3.0902e-01 12 8 8 10 0 0 99.96 18.01 0.29 -0.00245 -0.00216 11 11 1 5 0 2638.810 22.71 15.00 18.01 12.04 21.6 1033.24 -1014 -95 -532 -591 -328 467 55 701 -652 1635 -111 836 -533 -425 8 4 167 -38 166 -476 486 -259 325 286 1079 -493 886 85 -206 -326 -349 -514 82 161 -507 -167 -1132 492 -240 1145 -598 266 -262 2 -781 497 24 -26 257 -223 185 112 -49 -411 -68 75 118 -397 -531 -581 312 -167 -54 280 506 -32 108 464 -255 598 -604 89 -437 193 -282 111 -49 -140 -201 -229 420 -262 -290 161 386 -326 1 -365 973 -85 76 -153 414 3 -175 23 -614 379 -421 327 -552 642 -115 1204 88 879 435 264 347 357 516 -92 297 -201 88 90 284 70 198 -377 154 -592 -940 -300 -2 30 -204 -39 -141 -111 -288 -304 -419 247 -564 -65 -521 778 -201 195 107 -57 -310 -8 250 -192 367 -373 159 385 697 -73 494 -383 -155 -803 -372 -217 -569 323 -268 849 167 806 -337 178 -438 -123 240 -108 -416 37 -222 170 -131 -172 -30 26 104 -334 39 -812 -192 -136 211 -205 175 27 45 557 373 209 -9 393 -447 253 -178 48 -55 -185 -64 -60 -146 149 99 -322 123 -82 191 -108 23 -373 485 -30 391 147 96 64 -304 325 -333 405 147 518 -230 287 -191 219 -6 693 42 -117 -122 -255 -137 6 -46 -483 -236 -119 44 -685 278 -491 178 -472 -64 76 -422 -314 435 351 382 369 -43 309 -47 707 -21 666 -910 134 -42 28 -367 11 -510 -45 -249 -299 -179 270 33 425 -414 1077 -73 555 -366 -295 5 2 112 -25 109 -311 343 -199 266 234 907 -431 761 73 -177 -286 -311 -459 71 136 -427 -141 -953 423 -200 930 -499 225 -222 2 -677 449 22 -25 267 -235 198 121 -50 -425 -72 77 111 -346 -461 -508 280 -154 -49 251 457 -30 106 462 -251 561 -573 85 -429 190 -275 107 -47 -136 -186 -216 384 -244 -265 153 382 -337 1 -380 954 -84 76 -161 462 3 -191 22 -554 302 -347 302 -532 620 -105 1054 73 722 388 245 353 372 512 -87 274 -193 91 97 309 78 213 -388 153 -581 -962 -331 -3 34 -237 -44 -154 -124 -313 -325 -452 268 -601 -67 -523 770 -215 213 120 -63 -336 -9 275 -206 413 -432 183 434 763 -79 524 -414 -146 -722 -306 -179 -515 300 -276 941 182 881 -379 197 -484 -133 256 -108 -421 39 -229 184 -145 -194 -35 30 114 -359 39 -806 -198 -141 231 -223 185 29 49 607 396 211 -10 405 -445 250 -184 52 -64 -224 -80 -75 -179 180 116 -368 135 -90 204 -119 26 -440 581 -35 419 152 92 61 -310 361 -378 471 174 615 -278 338 -230 260 -7 829 50 -142 -151 -317 -170 7 -55 -557 -265 -132 49 -780 317 -582 207 -514 -69 74 -398 -298 391 323 383 376 -45 325 -45 628 -17 517 -712 108 -35 24 -349 11 SS 20231207T165623.7Z 8741.705830 411 1 29.88 2 66.47 177.03 72.24 245.05 0.00 1 0.00 1 0 33 1.464e-03 1.308e-03 1.267e-03 1.176e-03 4.418176e-07 3.812801e-07 5.897999e-08 3.559846e-07 4.581049e-07 222.30 -2.00 3.57 -2.00 340.40 0.00e+00 4.406523e-05 1.044974e-04 1.076992e-05 3.744714e-05 1.262488e-04 274.93 -2.00 20.00 -2.00 337.90 0.00e+00 -3.200089e-05 1.586825e-05 1.916080e-06 2.382211e-05 2.188752e-05 274.93 -2.00 20.00 -2.00 346.50 0.00e+00 6.164504e-04 4.548660e-05 3.495433e-05 1.533137e-05 6.130057e-05 274.93 -2.00 20.00 -2.00 345.11 0.00e+00 6.839326e+05 9.570509e+04 3.830589e+04 1.851777e+05 1.170471e+05 257.21 -2.00 11.30 -2.00 351.85 0.00e+00 3.011570e-01 1.487139e-03 2.552711e-03 3.087405e-03 1.538160e-03 222.30 -2.00 3.57 -2.00 330.58 0.00e+00 -2.980630e-04 1.928331e-03 4.501724e-04 1.013325e-04 1.630451e-03 274.93 -2.00 20.00 -2.00 328.61 0.00e+00 2.486 0.000 319.50 362.20 4 5.0226e+00 1.4970e-03 1.7292e-02 1.3569e-03 1.9575e-03 5.9219e-02 3.8657e-04 2.6981e-03 2.3992e-04 4.6693e-04 -8.0417e-03 7.4127e-04 5.1127e-04 7.3789e-05 9.0768e-04 3.1974e-04 3.7736e-04 4.4195e-04 2.6089e-05 3.6817e-04 7.2569e-04 5.9011e-04 3.2974e-04 5.2368e-05 6.6228e-04 1.1113e-03 4.2311e-03 9.0652e-04 2.1449e-04 5.3711e-05 1.0791e-03 -1.1240e-04 1.5424e-05 2.7583e-06 4.5911e-07 1.9851e-05 -1.2079e-03 0.0000e+00 3.5222e-01 9.1344e-02 1.7137e-02 2.7969e-03 1.1096e-01 12 9 8 11 8 0 99.98 18.02 0.17 -0.00121 -0.00260 27 27 1 5 0 1106.470 22.71 15.00 18.02 12.31 21.8 1035.37 74 -431 -636 -242 -255 -151 -66 -127 -4 242 85 258 445 449 -100 68 -156 -272 -130 -365 286 -33 161 294 79 305 22 -13 -24 23 143 10 56 -15 260 -11 -86 -41 -166 257 -54 -64 -115 56 -134 -95 -135 -228 -330 10 95 183 -66 51 -149 -243 -220 -238 -44 -55 5 18 72 154 -6 67 26 44 -157 58 -331 -2 -94 204 -107 217 -23 -14 105 -20 56 111 231 133 55 -37 -55 -93 -66 -142 177 -75 -86 -45 119 -94 -14 -241 45 70 129 25 84 209 117 403 4 86 -20 120 33 142 316 -68 36 -5 71 82 -272 -186 -96 -139 -167 0 70 112 151 -190 -27 93 -138 15 -157 -96 -192 -157 135 313 3 -86 19 87 47 -6 -145 -151 32 -4 25 -16 178 14 106 -40 80 196 -26 177 35 64 68 67 7 -54 -168 -61 -110 -162 -75 47 -94 -45 -174 -137 -85 -22 -39 -114 113 1 202 202 121 -81 117 55 -92 162 -96 192 -201 4 164 106 8 -1 -193 110 -148 16 103 82 77 -21 -40 -90 -164 -94 -53 -67 192 -72 139 -52 -41 -183 70 -56 -2 -187 -41 187 5 78 -27 240 32 19 -118 184 -93 24 -26 71 -60 -22 86 16 -70 -172 34 -69 -74 -249 201 -38 59 66 -51 -91 79 8 19 68 190 10 99 135 -5 31 -6 49 -101 -2 -72 -180 36 -211 -308 -118 -135 -85 -38 -75 -3 148 55 170 300 307 -67 43 -96 -162 -79 -238 202 -24 125 231 64 253 18 -11 -19 19 123 8 46 -12 210 -9 -70 -32 -133 202 -43 -53 -94 45 -105 -79 -121 -210 -321 10 95 174 -64 49 -148 -233 -194 -205 -37 -47 5 16 65 138 -5 60 24 42 -147 52 -298 -2 -88 192 -100 200 -21 -13 98 -18 49 98 214 123 53 -37 -54 -94 -64 -136 172 -77 -89 -48 121 -90 -12 -196 37 63 122 24 80 179 99 342 3 80 -20 119 33 135 290 -65 35 -5 74 88 -281 -192 -96 -136 -171 0 76 127 172 -214 -29 101 -146 16 -164 -101 -198 -156 138 318 3 -93 21 97 51 -7 -156 -161 36 -5 28 -19 196 15 113 -42 78 178 -21 148 33 63 70 71 7 -60 -192 -69 -123 -178 -81 47 -95 -48 -184 -152 -97 -26 -45 -129 128 1 201 200 126 -87 133 62 -100 179 -109 217 -223 4 170 106 8 -1 -208 119 -174 20 126 103 97 -25 -47 -105 -186 -103 -57 -74 216 -86 171 -62 -47 -191 70 -56 -2 -211 -48 225 6 93 -33 298 40 24 -149 223 -113 30 -33 89 -76 -27 105 19 -83 -202 41 -83 -87 -302 241 -44 67 67 -50 -89 78 8 20 75 211 10 100 125 -5 26 -5 43 -91 -1 -74 -198 SS 20231207T171224.4Z 8741.716949 416 1 3.20 2 66.42 181.06 74.88 248.05 0.00 1 0.00 1 0 33 1.665e-03 1.587e-03 1.591e-03 1.504e-03 -1.508034e-07 4.838301e-07 6.415363e-08 3.559832e-07 5.211116e-07 222.30 -2.00 3.57 -2.00 340.23 0.00e+00 4.342805e-04 1.337486e-04 1.196627e-05 3.744934e-05 1.442060e-04 274.93 -2.00 20.00 -2.00 337.71 0.00e+00 -3.665581e-05 2.024950e-05 2.344860e-06 2.382188e-05 2.478828e-05 274.93 -2.00 20.00 -2.00 346.39 0.00e+00 5.612865e-04 5.819283e-05 2.933293e-05 1.533323e-05 6.976407e-05 274.93 -2.00 20.00 -2.00 345.01 0.00e+00 3.151958e+05 1.217770e+05 3.720681e+04 1.851679e+05 1.340436e+05 257.21 -2.00 11.30 -2.00 352.01 0.00e+00 3.020791e-01 1.870040e-03 2.604701e-03 3.085988e-03 1.757625e-03 222.30 -2.00 3.57 -2.00 330.31 0.00e+00 -2.096696e-03 2.325456e-03 1.251430e-03 9.722038e-05 1.850788e-03 274.93 -2.00 20.00 -2.00 328.29 0.00e+00 2.482 0.000 319.50 362.20 4 3.2706e+00 1.9215e-03 1.7247e-02 1.3570e-03 2.2313e-03 2.8225e-02 4.9446e-04 2.6899e-03 2.4009e-04 5.3663e-04 -6.9054e-03 9.4717e-04 4.9539e-04 7.2968e-05 1.0396e-03 -2.0892e-04 4.6395e-04 4.6509e-04 2.6021e-05 4.2038e-04 1.7132e-03 7.4547e-04 3.4261e-04 5.1797e-05 7.6275e-04 6.4894e-03 1.4335e-03 1.1316e-03 2.4509e-04 5.6656e-05 1.2096e-03 5.3229e-05 1.9603e-05 2.8970e-06 4.4726e-07 2.2566e-05 -3.2903e-03 0.0000e+00 3.4823e-01 1.1560e-01 1.9417e-02 2.9364e-03 1.2586e-01 12 9 9 12 1 1 63.16 18.01 0.28 -0.00329 -0.00263 4 47 1 5 0 638.310 22.71 15.00 18.01 14.13 21.8 1046.18 -85 -347 -91 239 210 82 124 238 -49 21 -349 49 32 264 -355 226 -306 522 16 171 -299 134 -14 123 46 -102 33 81 -43 -244 57 -43 100 -164 138 -228 101 -170 -132 -174 108 -123 123 -71 -20 -90 52 -94 -52 102 246 101 -191 -33 -78 -12 49 27 154 -157 61 -475 28 253 448 -188 81 -4 -289 -416 170 235 282 -131 146 151 33 294 -228 206 221 130 146 -66 -180 47 -172 -79 -229 -61 -133 -199 274 -190 257 -8 -80 -312 -69 94 -81 211 190 -187 -416 392 -180 314 101 -58 -23 100 -86 65 88 143 -158 -253 205 -36 155 52 145 34 -132 -84 93 156 -59 -22 -245 -107 58 -209 -59 82 -66 165 -229 135 70 133 127 208 -55 63 -121 86 41 132 -136 -95 26 -131 57 -157 -66 -147 149 107 -23 -85 -105 146 -95 -170 -24 -48 10 -95 156 -99 -138 -13 97 -118 228 -49 -152 18 -62 207 -166 -31 -33 -118 -114 85 20 -116 83 -47 290 -131 44 -37 49 -47 103 -10 -252 210 15 -11 7 156 -51 353 159 52 -212 -23 77 -31 122 292 7 -11 -31 -25 108 -125 190 -109 -30 -87 -93 -255 227 295 -156 -49 -45 87 -296 49 -95 40 92 -19 -127 -96 -174 63 81 15 179 87 -14 -189 36 187 -4 -289 251 7 312 -270 -75 -167 -121 -137 -244 101 87 216 -49 -187 -48 137 126 50 76 149 -32 14 -240 34 22 186 -252 153 -207 339 10 117 -219 102 -11 99 38 -86 27 68 -35 -206 49 -37 83 -138 115 -193 84 -138 -106 -138 86 -101 101 -58 -16 -77 46 -88 -50 97 236 98 -188 -32 -75 -12 45 24 133 -137 54 -435 25 224 400 -171 76 -3 -269 -383 155 218 262 -124 139 143 31 280 -208 183 196 114 135 -63 -173 46 -170 -78 -222 -59 -130 -200 281 -200 263 -8 -72 -262 -58 85 -77 204 176 -166 -356 334 -163 300 102 -59 -23 95 -82 62 88 148 -164 -268 209 -36 152 51 146 36 -144 -94 103 171 -64 -23 -257 -110 60 -220 -61 83 -65 163 -240 147 77 144 137 223 -58 68 -135 96 45 146 -149 -103 28 -138 56 -143 -58 -130 137 104 -23 -91 -115 161 -105 -191 -26 -52 10 -96 159 -104 -147 -14 107 -133 259 -56 -167 19 -63 208 -172 -33 -36 -130 -124 93 22 -129 89 -51 299 -135 46 -38 52 -52 118 -12 -307 253 19 -13 8 183 -58 392 171 58 -241 -27 92 -36 137 312 7 -11 -33 -28 125 -148 222 -129 -36 -107 -113 -309 275 360 -191 -60 -56 108 -363 60 -114 47 108 -22 -148 -112 -206 76 96 17 198 90 -15 -189 36 190 -4 -322 281 7 324 -264 -69 -152 -109 -125 -225 99 93 240 SS 20231207T171712.0Z 8741.720278 419 1 29.66 2 66.45 182.27 75.68 248.93 0.00 1 0.00 1 0 33 1.037e-03 9.241e-04 7.736e-04 7.231e-04 8.905481e-07 2.336087e-07 5.483699e-08 3.559819e-07 3.247407e-07 222.30 -2.00 3.57 -2.00 340.37 0.00e+00 1.432937e-04 6.444696e-05 1.407159e-05 3.744861e-05 8.974126e-05 274.93 -2.00 20.00 -2.00 337.89 0.00e+00 1.509403e-05 9.725068e-06 2.090608e-06 2.382206e-05 1.546705e-05 274.93 -2.00 20.00 -2.00 346.49 0.00e+00 5.401456e-04 2.794901e-05 2.972174e-05 1.532794e-05 4.346454e-05 274.93 -2.00 20.00 -2.00 345.10 0.00e+00 4.101136e+05 5.870654e+04 3.647083e+04 1.851747e+05 8.327566e+04 257.21 -2.00 11.30 -2.00 351.89 0.00e+00 3.007927e-01 9.135448e-04 2.587463e-03 3.086178e-03 1.092392e-03 222.30 -2.00 3.57 -2.00 330.48 0.00e+00 -2.901489e-03 1.163215e-03 1.452164e-03 9.976231e-05 1.154138e-03 274.93 -2.00 20.00 -2.00 328.48 0.00e+00 2.484 0.000 319.50 362.20 4 4.1486e+00 9.2114e-04 1.7265e-02 1.3570e-03 1.3870e-03 4.2555e-02 2.3793e-04 2.6943e-03 2.4001e-04 3.3251e-04 -5.6588e-03 4.5552e-04 4.9541e-04 7.3525e-05 6.4521e-04 3.4357e-06 2.2941e-04 4.6728e-04 2.5891e-05 2.6151e-04 1.2156e-03 3.6197e-04 3.4055e-04 5.2291e-05 4.7265e-04 2.6718e-03 5.6980e-03 5.5184e-04 2.2244e-04 5.4357e-05 7.5484e-04 -2.1496e-05 9.4472e-06 2.8523e-06 4.5764e-07 1.4064e-05 -2.4475e-03 0.0000e+00 3.1988e-01 5.5894e-02 1.7480e-02 2.8797e-03 7.8532e-02 12 8 8 10 0 0 99.93 18.02 0.22 -0.00245 -0.00271 37 37 1 5 0 801.530 22.71 14.99 18.02 14.31 21.7 1046.03 -212 -506 -71 -416 -22 14 -189 -77 -86 210 95 294 61 371 89 -18 -163 -103 27 -1 157 210 131 202 -8 -17 46 13 154 -76 -66 -72 -78 40 48 4 -27 -37 -120 -95 -42 81 -7 162 -13 -35 11 110 -20 -64 -148 64 -115 -38 -86 -122 -29 -136 144 134 6 60 41 61 -59 33 -22 -60 -164 70 -84 -66 45 73 85 -42 -84 -74 -96 -127 97 -92 -1 -79 133 -128 33 -48 20 -206 -188 -110 153 12 14 72 -30 215 194 82 102 47 125 144 25 98 -1 219 83 87 -31 37 52 48 -51 53 -35 -38 19 -162 41 1 -103 78 100 5 56 147 6 34 -19 -196 -64 -67 -140 -64 -59 1 -45 42 -67 -19 -43 -67 -53 -67 -40 -45 -71 -78 164 7 37 37 129 76 49 11 17 86 -110 34 -19 85 28 -60 5 64 -132 103 -63 -66 49 -55 -54 112 22 -57 17 16 6 18 -53 -209 -52 55 47 -35 -20 26 -142 -38 80 67 -33 -122 3 -13 3 153 181 69 58 18 -43 16 -56 57 -4 -108 138 -79 138 -4 45 25 -92 55 -31 -48 -107 33 -63 25 -143 -21 -15 20 -45 136 -11 131 53 143 99 3 -53 -99 98 -125 -137 -117 24 -4 -12 -159 -61 43 22 82 -24 -104 -114 -36 -73 105 42 116 -15 126 184 -16 119 -20 -63 -9 -110 -263 -36 -216 -13 8 -110 -46 -53 137 62 197 42 257 60 -12 -108 -66 18 -1 113 154 98 163 -7 -14 38 11 130 -63 -55 -60 -63 34 40 3 -22 -30 -93 -73 -33 65 -6 128 -11 -30 9 101 -20 -62 -140 61 -112 -37 -81 -117 -26 -115 122 114 5 54 36 54 -53 30 -20 -57 -153 66 -79 -60 42 69 79 -38 -78 -69 -90 -113 84 -81 -1 -73 128 -128 33 -47 20 -198 -183 -108 161 13 14 70 -26 172 159 73 97 45 118 124 21 83 -1 203 82 88 -31 36 48 46 -51 55 -37 -39 20 -161 42 1 -102 81 111 6 63 161 7 35 -21 -206 -67 -69 -147 -65 -59 1 -48 46 -74 -21 -48 -72 -56 -74 -44 -51 -81 -87 178 8 41 38 127 70 43 10 16 85 -115 37 -20 95 31 -67 6 69 -139 106 -66 -69 54 -61 -61 129 25 -65 19 17 6 19 -54 -223 -59 62 51 -39 -23 30 -157 -40 82 70 -35 -130 3 -14 4 183 225 85 71 21 -52 18 -64 63 -4 -120 156 -93 167 -5 50 27 -93 55 -33 -54 -126 39 -76 31 -175 -25 -19 24 -55 170 -14 161 66 178 124 3 -64 -118 115 -144 -160 -139 29 -5 -14 -182 -67 45 22 80 -25 -107 -123 -41 -80 115 44 109 -14 112 162 -14 107 -19 -66 -10 SS 20231207T172225.9Z 8741.723911 422 1 31.08 2 66.50 183.58 76.56 249.88 0.00 1 0.00 1 0 33 1.538e-03 1.439e-03 1.449e-03 1.359e-03 7.628102e-07 4.373819e-07 7.000900e-08 3.559897e-07 4.813059e-07 222.30 -2.00 3.57 -2.00 340.18 0.00e+00 1.345845e-05 1.197943e-04 1.572833e-05 3.744958e-05 1.321266e-04 274.93 -2.00 20.00 -2.00 337.67 0.00e+00 8.853591e-06 1.837371e-05 2.399850e-06 2.382172e-05 2.304626e-05 274.93 -2.00 20.00 -2.00 346.33 0.00e+00 5.435760e-04 5.278425e-05 2.568164e-05 1.533697e-05 6.440051e-05 274.93 -2.00 20.00 -2.00 344.89 0.00e+00 1.480560e+05 1.100443e+05 3.598015e+04 1.851839e+05 1.228317e+05 257.21 -2.00 11.30 -2.00 351.74 0.00e+00 3.018206e-01 1.684248e-03 2.561635e-03 3.086871e-03 1.614654e-03 222.30 -2.00 3.57 -2.00 330.43 0.00e+00 2.114868e-03 2.154513e-03 5.493558e-04 1.001155e-04 1.712850e-03 274.93 -2.00 20.00 -2.00 328.43 0.00e+00 2.489 0.000 319.50 362.20 4 5.4052e+00 1.7307e-03 1.7292e-02 1.3567e-03 2.0592e-03 7.6504e-02 4.3988e-04 2.6983e-03 2.3994e-04 4.8759e-04 -1.3550e-02 8.5209e-04 4.9299e-04 7.3199e-05 9.5427e-04 1.7511e-04 4.2550e-04 4.4254e-04 2.6055e-05 3.8611e-04 3.4575e-03 6.7295e-04 3.1972e-04 5.1630e-05 6.9421e-04 7.6085e-04 -1.5187e-03 1.0403e-03 2.1810e-04 5.5338e-05 1.1466e-03 -9.7750e-05 1.7849e-05 2.8630e-06 4.5760e-07 2.0860e-05 -1.4795e-03 0.0000e+00 4.1233e-01 1.0522e-01 1.7664e-02 2.8463e-03 1.1660e-01 12 9 8 11 8 0 99.59 18.02 0.35 -0.00148 -0.00269 10 10 1 5 0 3107.510 22.71 15.00 18.02 14.35 21.8 1046.19 59 -565 -309 171 320 -295 213 -202 -128 118 -12 34 83 152 111 -114 -110 -61 -110 -214 60 350 261 425 209 -123 -185 6 -305 -148 -76 -33 263 -41 20 64 -124 167 209 151 -253 96 -50 -196 149 -235 -46 -121 143 76 -25 84 154 115 -87 124 56 -67 -92 -251 -229 -249 122 -33 -54 -51 128 9 298 -175 -145 53 27 -184 -78 -64 153 -280 57 -8 -163 -91 -95 -211 -109 -115 95 -170 170 38 27 55 -122 -33 120 -39 -43 102 318 232 50 9 243 40 386 376 37 366 266 36 74 18 31 89 -173 119 293 -138 -80 129 44 -76 -102 -288 -229 194 -39 -87 -147 -152 -256 97 106 -54 -205 21 -176 10 -53 -77 147 105 1 -51 16 -174 -8 27 -120 -55 160 -66 142 -116 5 158 -346 136 -123 125 -52 -30 -73 310 136 70 -16 0 -204 -65 -87 -29 34 -64 -99 -147 59 -30 -90 -52 -154 36 -70 9 59 124 -20 60 136 -78 165 137 -88 69 -32 -220 67 -7 153 -133 159 21 120 -14 171 -88 -9 -130 -117 124 76 -61 184 269 -12 -141 178 -76 -91 -80 -134 -101 228 99 -26 -89 25 149 42 117 81 -83 -12 110 -5 86 -7 -331 -192 -271 -204 -81 103 42 13 -72 -143 117 16 78 138 31 -232 -321 81 64 83 217 -71 116 9 40 -68 105 111 41 31 -286 -154 91 173 -172 128 -123 -81 74 -8 23 58 107 76 -75 -70 -40 -74 -143 44 258 207 341 174 -106 -158 5 -258 -127 -67 -29 223 -34 16 53 -102 133 171 122 -206 80 -41 -162 125 -198 -42 -113 141 76 -24 81 152 114 -84 121 50 -59 -80 -221 -209 -234 113 -30 -50 -46 121 9 281 -160 -132 49 26 -177 -73 -60 145 -264 54 -7 -144 -81 -89 -198 -106 -116 94 -174 165 36 26 54 -128 -35 123 -38 -38 84 267 209 48 9 232 35 329 322 32 340 272 36 74 18 29 85 -172 122 305 -148 -83 133 44 -75 -105 -306 -250 220 -44 -98 -159 -165 -270 100 111 -57 -211 20 -180 10 -55 -84 164 116 1 -56 17 -187 -9 31 -138 -62 176 -71 152 -124 4 144 -289 114 -113 121 -53 -32 -80 343 154 78 -17 0 -218 -65 -88 -30 35 -70 -112 -169 67 -33 -102 -53 -153 35 -72 9 67 139 -21 65 153 -88 181 142 -91 72 -32 -224 72 -7 177 -161 193 25 148 -17 200 -101 -10 -142 -124 135 89 -73 224 317 -14 -152 178 -75 -96 -89 -155 -120 263 115 -31 -109 31 184 52 146 100 -101 -15 136 -6 105 -8 -397 -225 -314 -237 -95 123 50 15 -82 -159 123 16 75 135 31 -243 -347 88 67 83 198 -63 97 7 34 -60 98 111 44